LPA-1 Specifications

Performance Specifications

    Precision of Results
    Wood, Drywall ±0.05 mg/cm2
    Concrete, Metal ±0.10 mg/cm2

    Accuracy of Results
    Wood, Drywall ±0.05 mg/cm2
    Concrete, Metal ±0.10 mg/cm2

Instrument Characteristics

    Depth of PenetrationNo Interference from objects (lead pipes, etc.) beyond 3/8"
    Substrate EffectAutomatic Compensation (Spectrum Analysis)
    Detection Technique X-Ray Fluorescence (XRF)
    Weight<3lb. Total (one piece unit includes 1lb battery pack)
    ErgonomicGrip-handle for ease of measurements and limited operator fatigue.
    Measurement AreaThe face print is smaller than 1.5" X 1.75".
    System battery life8 hours of normal usage between charges for each battery
    Battery Charge Time 1.5 hours
    Operating Temperature20-120°F

Radiation Safety Features

    Manual Shutter LockThe shutter will not open when the manual shutter lock is activated.
    Trigger LockA key activated trigger lock renders the trigger inoperable.
    Optical SensorsEnsure that the bezel of the instrument is flat against the surface before the shutter can be opened.

Radiation Source

    Cobalt-57 (Min. 10 mCi, Max. 12 mCi, 444 MBq)
    RMD's LPA-1 analyzer radioactive source is a 12 mCi, Cobalt 57 isotope. Co-57 gamma ray energies (122 Kev) are more efficient (at least 10 times) than those of Cd-109 (22-26 Kev and 88 Kev) source for production of K-shell lead x-rays, which are recommended by HUD chapter 7 for measurement of lead in paint. Therefore, the measurement time is shorter using a Co-57 source

Detector

    The LPA-1 analyzer utilizes a "room-temperature" detector, CdTe, for its x-ray analysis that operates in a wide, 20-120° F, thermal range without any cooling requirements. The LPA-1 analyzer's ability to operate without cooling eliminates a cooling down period for stability at the beginning of a job that would result in extra time added to inspections.

Operation

    The LPA-1 XRF system has the capability of setting abatement levels from 0.4-2.0 mg/cm2 for automatic determination of the lead condition as either positive or negative with 95% confidence according to a PCS approved mode of operation.

Memory

    The LPA-1 can data log up to 4000 data points and is easily down loaded through an RS-232 serial interface port for report generation with the WindowsTM based Report Generation Software.

Software

    The WindowTM based interactive graphic software is considered to the most comprehensive, universally applicable lead paint report generation system available. The software is included in the LPA-1 instrument package and upgrades are free for as long as you own the instrument.

Warranty

    2-year parts and labor warranty with a life time warranty on the calibration of the instrument. RMD Instruments Corp also offer upgrades of the software and system upgrades free of charge for as long as you own your instrument.
All claims of performance are verified by the Performance Characteristic Sheet, Edition No.: 4, developed by QuanTech, Inc., under a contract from the U.S. Department of Housing and Urban Development (HUD).

Accessories

    The LPA-1 Inspection System includes everything necessary to be a fully functional Lead Paint Inspector, including:
    Two rechargeable batteries8 hours of normal usage between charges for each battery.
    One battery chargerQuick Charge time of 1.5 hours.
    Set of two stabilizersStabilizes the instrument's small measurement surface on flat surfaces.
    One interface cableAllows for download from RS 232 to Serial Port.
    One carrying caseLightweight, waterproof metal case with two key locks and comfortable leather handle. The case measures 19" x 15" x 9"
    One calibration standard1.0 mg/cm2 N.I.S.T. traceable wood standard.


Links   |   Site Map
RMD Instruments Corp   44 Hunt Street   Watertown, MA 02472
phone: (617) 668-6901   fax: (617) 926-9743 email: LPA@rmdinc.com