LPA-1 OperationOver ten years ago a group of experienced Lead Inspectors came to RMD in need of a better XRF lead paint spectrum analyzer. With their input, RMD developed the easy to use LPA-1. Operating the LPA-1 in the field to complete a Lead Paint inspection involves 4 simple steps. 1. Choose a Job Number
A job number will appear which is created from the date and military time at which the inspection began. For Example a Job Number 12-15-03-0945 (12/15/03 09:45) indicates that the inspection began on December 15, 2003 at 9:45am. Both the Analyzer and the Report Generation Software use this number to identify inspection jobs.
2. Choose an Abatement Level
Abatement Level can be set in increments of 0.1 mg/cm2 to a maximum of 2.0 mg/cm2. In Quick Mode, the Analyzer uses the abatement level to determine the measurement time required for a 95% confidence. 3. Confirm Calibration In TIME CORRECTED mode (SELECT MODE key), measure the Standard Sample (1.0mg/cm2) a minimum of 3 times. Avg. of 3 readings should equal 1.0 ± 0.3 (0.7-1.3) mg/cm2 Also, In TIME CORRECTED mode measure the blank side of the Standard Sample 3 times for a 0.0mg/cm2 reference.
4. Start Inspection
Repeat Step 3 before ending the job. Step 3 should also be done every 4 hours within the same job number. |
|